Facebook Twitter LinkedIn YouTube Menu Search Arrow Right Arrow Left Arrow Down Arrow Up Home Arrow Next Arrow Previous RSS Icon Calendar Icon Warning Icon

Articles in Industry Publications

Article

Considerations for measuring & test equipment (M&TE) resolution

  • November 2018
  • Number of views: 7703
  • Article rating:

By Mike Howell
EASA Technical Support Specialist

Accuracy and precision of measuring & test equipment (M&TE) have been topics of previous Currents articles (November and December 2014). A related topic that was not covered in the previous articles is resolution. The Joint Committee for Guides in Metrology document JCGM 200:2012 defines resolution as “the smallest change in a quantity being measured that causes a perceptible change in the corresponding indication.” Simplified, it’s the smallest difference that can be measured by the subject equipment. The accuracy of the M&TE must be greater than (less accurate) or equal to the resolution. That is, the M&TE must be able to indicate the value that is compared to the standard during calibration.

LOGIN TO READ MORE

AVAILABLE IN SPANISH

Documents to download



Comments are only visible to subscribers.