Considerations for measuring & test equipment (M&TE) resolution

  • November 2018
  • Number of views: 7885
  • Article rating:

By Mike Howell
EASA Technical Support Specialist

Accuracy and precision of measuring & test equipment (M&TE) have been topics of previous Currents articles (November and December 2014). A related topic that was not covered in the previous articles is resolution. The Joint Committee for Guides in Metrology document JCGM 200:2012 defines resolution as “the smallest change in a quantity being measured that causes a perceptible change in the corresponding indication.” Simplified, it’s the smallest difference that can be measured by the subject equipment. The accuracy of the M&TE must be greater than (less accurate) or equal to the resolution. That is, the M&TE must be able to indicate the value that is compared to the standard during calibration.